Get in Touch
Follow us

Buraq Integrated Solutions

Founded in 2000, Buraq Integrated Solutions (BIS) has successfully executed multidisciplinary projects across Bangladesh, Maldives, Singapore, Thailand, Indonesia, Turkey, UAE, Bahrain, KSA, Australia, and the USA.

Contact Us

Location
HQ: Buraq Integrated Solutions,
11-D, 6th Road, Satellite Town, Rawalpindi, Pakistan.

STZ: Buraq Integrated Solutions,
1st Floor, Alpha-18, NASTP, Old Airport Road, Rawalpindi, Pakistan.

AUS: Buraq Integrated Solutions,
Unit 55, 10 Yato Road, Prestons NSW 2170, Australia.

Phone
+92 51 484 0137
+92 333 513 2514
+92 300 512 1896
Email
Introduction to Sensor

Conductivity Plus

Accurate Conductivity, Water Level & Temperature Monitoring in One Instrument

The Conductivity Plus is a precision monitoring instrument designed to measure electrical conductivity, water levels, and temperature in wells, boreholes, standpipes, and estuaries. It is ideal for groundwater investigations, saltwater intrusion monitoring, and contamination assessment, helping detect changes in water quality through variations in conductivity and salinity.

Specifications

Specifications

Reel Assembly

Specification

Standard

Large

Height

36cm (14″)

43cm (17″)

Depth

21cm (8.25″)

23cm (9″)

Width

29cm (11″)

34cm (13.5″)

Weight (Reel Assembly Only)

2.60kg (5.7lb)

4.2kg (9.3lbs)

Plate Size

27cm (10.5″)

34cm (13.5″)

Plate Material

Nylon Fibre Composite

Aluminum

Brake

Ergonomic Dial

Ergonomic Dial

Stainless Ball Bearing

YES

YES

Hanger

Included on frame

Included on frame

Tape Guide

Included on frame

Included on frame

Field Testing

Included Test Kit

Included Test Kit

Tape Lengths

30m, 100ft

200m, 750ft

 

60m, 200ft

300m, 1000ft

 

100m, 300ft

Ask about lengths up to 5000ft

 

150m, 500ft

 

Tape

Specification

Details

Tape

High Tensile Stainless Steel

Jacketing

Polyethylene

Conductors

7 Strand (4 SS316 & 3 Copper)

Markings

Metric or Engineering Scale – Under

Width

9.5mm (3/8″)

Dog Bone Profile

Dogbone for low friction

Break Strength

125kg (280lbs)

Accuracy Certification

ASME B89.1.7-2009, MIL-STD-45662A, ISO 10012-2003

Accuracy Compliance

FED GGG-T-106F, EEC Class II, USGG-T-

Break Strength

127kg (280lbs)

Electronic Panel

Specification

Details

Field Replaceable

Yes (with philips #2 screwdriver)

Screen

LCD

Water resistance

IP67

Temperature Units

°C, °F, °K, °R

Conductivity Unit

µS/cm

Battery Indicator

On startup – LCD display and audio indication

Visual Indications

LED Light and LCD Display

Audible Indications

Volume adjustable buzzer, Silence button

Buttons

Button 1 (Power/hold for Units), Button 2 (Silence/hold to Calibrate)

Link & Probe

Specification

Details

Link & Probe

Snag Free Design

IP Rating

IP68

Weight

125g (4.4oz)

Length

16mm

Diameter

16mm (5/8″)

Maximum Recommended Probe

70°C (160°F)

Removable

Yes- Field Replaceable

Depth Rating

1.5km (5,000ft)

Break Strain

Yes 70kg (150lbs)

Probe Wetted Materials

Stainless Steel 316, Delrin, Gold, Buna, PFAS Free

Standard Link Wetted Materials

Stainless Steel 316, Delrin, Polyurethane, Buna, PFAS Free

PFAS Status

Manufactured with wetted materials not known to contain PFAS

Conductivity Accuracy

Conductivity Accuracy

Test Condition Range

Typical

Max Deviation

Calibration Solution

Conditional upon calibrating to all 4 calibration points

10-3200µS/cm

+/-1%

+/-2%

1413µS/cm

 

3207-8840µS/cm

+/-1.25%

+/-2%

5000µS/cm

 

8941-62340µS/cm

+/-1.25%

+/-2%

12880µS/cm

 

62341-250000µS/cm

+/-1.5%

+/-3%

111800µS/cm

Additional Conductivity Specifications

Specification

Details

Conductivity Repeatability

8 Averages

Typical

+/-0.06%

Max Deviation

+/-0.2%

Conductivity Resolution

1µS/cm

Conductivity Response

<1 second (After temperature stabilization)

Recommended Conductivity Reading

10-250000µS/cm

Lifetime Conductivity Stability

Dependent on usage. For best results, recalibrate often.

Total Conductivity Reading Range

10-999999µS/cm

Temperature Compensation

Temperature compensated to be expressed at international standard of 25°C

Calibration Range Solutions

1413µS/cm, 5000µS/cm, 12880µS/cm, 111800µS/cm

Temperature Compensation

2%/°C

Conductivity Calibration Point

Closest to Reading

Temperature Specifications

Specification

Test Condition

Typical

Max

Temperature Accuracy Typical

-20°C to +50°C

+/-0.05°C

+/-0.1°C

Temperature Resolution

0.0078°C

  

Temperature Reading Range

-40°C to +70°C

  

Temperature compensated range

-20°C to +70°C

  

Temperature Response Time

10 seconds per degree Celsius

  

Additional Temperature Specifications

Specification

Details

Value

Lifetime Temperature Stability

300 hours continual reading at 70°C

+/-0.03°C

Temperature

8 Averages

+/-0.0078°C

Temperature cycling and hysteresis

8 Averages

+/-0.0078°C

Features

Special Features

01

Narrow 16mm Probe Design

The compact 16mm (5/8”) diameter probe allows convenient deployment in wells, boreholes, and standpipes.
02

Automatic Calibration Detection

Automatically identifies the calibration solution, simplifying calibration and saving time during field operations.
03

Triple-Sealed Removable Probe

The electronic probe features triple sealing and a removable design for enhanced protection and convenient maintenance.
04

Complete Calibration Kit Included

Designed to provide best-in-class battery life for prolonged field operation and reduced battery replacement frequency.
Applications

Primary Applications

borehole
Borehole Profiling
Environmental Monitoring
Hydrological Monitoring
water management
Water Management