Introduction to Sensor
Digital Output Quantum Sensor (SQ-421X-SS)
Reliable PPFD Measurement for Agricultural and Aquatic Environments
The SQ-421X-SS Quantum Sensor is a digital sensor designed for accurate PPFD (Photosynthetic Photon Flux Density) and PAR measurements using the SDI-12 communication protocol. Calibrated for sunlight, it includes automatic immersion correction for underwater measurements. Its rugged, self-cleaning housing and reliable cable design make it suitable for outdoor fields, greenhouses, growth chambers, plant canopies, and aquatic environments.
Specifications
Specifications
| Parameter | Specification |
|---|---|
| Input Voltage Requirement | 5.5 to 24 V DC |
| Current Draw | 1.4 mA (quiescent); 1.8 mA (active) |
| Calibration Uncertainty | ±5% |
| Measurement Repeatability | Less than 1% |
| Long-Term Drift (Non-Stability) | Less than 2% per year |
| Non-Linearity | Less than 1% (up to 2500 µmol m⁻² s⁻¹) |
| Response Time | 0.6 s; detector signal reaches 95% following a step change. Fastest SDI-12 data transmission rate: 1 s |
| Field of View | 180° |
| Spectral Range | 370 to 650 nm (wavelengths where response is greater than 50% of maximum) |
| Directional (Cosine) Response | ±5% at 75° zenith angle |
| Temperature Response | Less than 0.5% from −20 to 50 °C |
| Operating Environment | −20 to 60 °C; 0 to 100% relative humidity; submersible up to 30 m depth |
| Dimensions | 30.5 mm diameter × 37 mm height |
| Mass | 140 g (with 5 m cable) |
| Cable | 5 m two-conductor, shielded, twisted-pair wire; additional cable available in 5 m increments; TPR jacket; pigtail lead wires |
| Warranty | 4 years against defects in materials and workmanship |
| Manufactured | Assembled in the USA |
Features
Special Features
01
Accurate PAR Measurement
Provides accurate and stable measurements of photosynthetically active radiation for agricultural and research applications.
02
Full-Spectrum and Original Models
Available in Full-Spectrum Quantum and Original Quantum versions to suit different light sources and measurement requirements.
03
NIST-Traceable Calibration
Calibration is performed using reference standards with traceability to the National Institute of Standards and Technology.
04
Wide Light Source Compatibility
Original models are suitable for broadband light sources, while Full-Spectrum models support all light sources, including LEDs.
Applications